VIVAFLOW 200, 3,000 MWCO PES, 2/PKG
0.5 - 5 l
- Plug and play laboratory cross flow cassette
- Re-usable
- Optimal for cell culture supernatant and virus concentration
- High flow rate and recovery
-
Vivaflow® 200 is a ready to use and reusable laboratory crossflow cassette for concentrating up to 5 l samples. The membrane surface per device is 200 cm². One unit comes with all necessary accessories for running the device with a laboratory pump and a size 16 pump head. For speeding up the concentration, two cassettes can be run simultaneously.
Vivaflow® 200 is optimally suited for applications like cell culture supernatant concentration, virus concentration and water concentration e.g. in fish breeding facilities. This economical product offers a standard of ease of use, reliability and flexibility unrivalled by any laboratory filtration system in the market. The thin-channel, flip-flow recirculation path of the modules provides high crossflow velocities with minimum pump speed requirements. A single 200 cm2 module typically concentrates 1 l by a factor of 50 in less than 30 min. Alternatively, two Vivaflow® 200 modules can be used in parallel for concentration of 5 l in under 75 minutes. Near total recovery of the concentrate is achieved with a single final rinse. Vivaflow® 200 can be easily cleaned and re-used several times. All accessories for sample concentration, except a peristaltic pump with a pump head for size 16 tubing, are included in the package for using a single cassette. A 500 ml sealed diafiltration reservoir can be added to the system for rapid and convenient buffer exchange.
Equipment required
The standard Vivaflow® 200 package comes with tubing, pressure indicator, flow restrictor and high pressure pump tubing. All you need is a peristaltic pump with a pump head capable of handling 6.4 mm OD tubing (size 16).
Equipment required
The standard Vivaflow® 200 package comes with tubing, pressure indicator, flow restrictor and high pressure pump tubing. All you need is a peristaltic pump with a pump head capable of handling 6.4 mm OD tubing (size 16).
Running 2 Vivaflow® 200 cassettes in parallel
For speeding up the concentration, two Vivaflow® 200 cassettes can easily be connected using a Y connector which is preassembeled with the correct tubing sizes and fittings (size 16 tubing for connecting the two Vivaflow® 200 units, size 15 tubing for the pump). Please note that you need a size 15 pump head for running 2 Vivaflows® in parallel.
- High flow rate and recovery
- Clear housing for process monitoring
- Ready to use single cassettes
- Easy set up of 2 cassettes for use in parallel
- Re-usable product
Performance Characteristics
Time to concentrate up to 20x (min.) at 3 bar
inlet pressure, 20°CStart volume Average flux Recovery % 1 litre ml/min Direct 25 ml rinse Lysozyme 0.25 mg/ml (14,000 MW) 2,000 MWCO Hydrosart 160 6 97 % >99 % 3,000 MWCO PES 180 5 97 % >99 % BSA 1.0 mg/ml (66,000 MW) 5,000 MWCO PES 29 33 98 % >99 % 5,000 MWCO Hydrosart 70 14 98 % >99 % 10,000 MWCO PES 23 41 96 % >99 % 10,000 MWCO Hydrosart 35 27 98 % >99 % 30,000 MWCO PES 25 38 96 % 99 % 30,000 MWCO Hydrosart 20 48 96 % >99 % 50,000 MWCO PES 22 43 96 % 98 % γ Globulins 1.0 mg/ml (average 160,000 MW) 100,000 MWCO PES 54 18 96 % 99 % 100,000 MWCO RC 45 21 96 % 99 % Yeast 1.0 mg/ml (S. Cerevisiae) 0.2 μm PES 11 86 92 % 98 % Dilute solute concentration, start volume 1 litre at 3 bar, 10,000 MWCO PES BSA 0.001 mg/ml 18 52 90 % 98 % BSA 0.01 mg/ml 20 47 92 % 98 % BSA 0.1 mg/ml 21 45 94 % 99 % Start volume 5 litres (two VF200 in parallel at 3 bar) 10,000 MWCO PES BSA 1.0 mg/ml (66,000 MW) 67 70 97 % >99 %
| Product properties | |
| MWCO | 3,000 Daltons |
| Width | 38 mm |
| Volume Range | 500 – 5,000 mL |
| Max. operating pressure | 4 bar |
| Primary Product Type | Ultrafiltration Devices |
| Membrane material | Polyethersulfone (PES) |
| Membrane | Polyethersulfone (PES) |
| Tubing | PVC (medical grade) |
| Material o-ring | seal | Silicone (SI) |
| Technical attributes | |
| Electrical / battery item | false |
| Housing material | Acrylic |
| Pressure indicator | Polypropylene, SS spring |
| Flow Channel | Acrylic |
| Flow Restrictor | Polypropylene (PP) |
| Maximum Pressure | 4 bar |
| Overall length | 127 mm |
| Hold-up volume | 5.3 mL |
| Active Membrane Area | 200 cm² |
| Overall height | 138 mm |
| Max. temperature | 60° C |
| Non-recoverable Hold-up | < 2 mL |
| Channel width | 10 mm |
| Minimum Recirculation Volume | < 20 mL |
| Channel height | 0.4 mm |
| Pump Flow | 200 – 400 mL/min |
| Application | Protein Concentration |
| Area of application | Buffer Exchange` Protein/DNA Concentration |
| MarketResearch | Research |
| Device typeTangential / Crossflow Concentrator | Tangential / Crossflow Concentrator |
| Sample TypeProtein` Peptide` Nanoparticles | Protein` Peptide` Nanoparticles |
| FittingsNylon (NY) | Nylon (NY) |
| Membrane support | Polypropylene (PP) |